Time-of-flight–resolved interferometric speckle-contrast optical spectroscopy (TOF-iSCOS) for depth-resolved blood-flow sensing

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Significance

Continuous, noninvasive, and depth-resolved monitoring of blood-flow-related tissue dynamics remains an important unmet need. Speckle-contrast optical spectroscopy (SCOS), including interferometric implementations such as iSCOS, provides a scalable optical route to blood-flow sensing, but conventional continuous-wave approaches lack intrinsic depth selectivity. Time-of-flight (TOF) gating offers a way to separate superficial and deeper dynamic contributions in layered tissues, such as skin– muscle or scalp–cortex, by resolving photon path lengths.

Aim

We introduce a swept-source, single-channel implementation of interferometric speckle-contrast optical spectroscopy (iSCOS) to obtain TOF-resolved temporal speckle contrast, κ 2 , from the measured field autocorrelation g 1 , and evaluate its feasibility for depth-resolved blood-flow sensing.

Approach

A swept-source iNIRS system operating at 780 nm acquired interferometric signals, which were Fourier-transformed along the optical-frequency axis to recover complex TOF-resolved speckle fields. Temporal speckle contrast was then estimated at each TOF gate indirectly from g 1 using the speckle-visibility relation. Diffusion-based numerical simulations were first used to compare the direct variance-based estimator and the indirect g 1 -based estimator under varying reduced scattering coefficient, diffusion coefficient, additive noise level, and bi-layer geometry. Because the simulations showed that the g 1 -derived κ 2 estimator was substantially less sensitive to additive noise than the direct estimator, this estimator was used for the main phantom and in vivo analyses, while the direct estimator served as a simulation comparator. The g 1 -derived estimator was then applied to liquid and bi-layer phantoms, followed by proof-of-concept in vivo measurements on the human forearm during cuff occlusion and on the forehead during a Sudoku task.

Results

TOF-resolved κ 2 curves recovered with the g 1 -derived estimator matched DWS theory across scattering coefficients, photon path lengths, and exposure times. The estimator preserved theoretical accuracy for additive noise amplitudes up to 50% of the field amplitude, whereas the direct variance estimator showed substantial noise-induced bias and required correction. Bi-layer simulations and phantom experiments reproduced the predicted direction and onset of TOF-dependent decorrelation-rate trends in layered media. In vivo, the recovered blood-flow index tracked the expected TOF-dependent cuff-occlusion and reactive-hyperemia response in the forearm. During the single-subject Sudoku task, the left-forehead recording showed a TOF-dependent relative blood-flow-index increase of +0.8 ± 1.9% at TOF = 400 ps, +9.8 ± 2.2% at TOF = 600 ps, and +15.2 ± 5.6% at TOF = 800 ps. This pattern is consistent with increased sensitivity to deeper tissue at longer photon path lengths, but requires cohort-level validation before quantitative interpretation as cognitive activation.

Conclusions

Coupling temporal speckle-contrast analysis with swept-source iNIRS yields a proof-of-concept, depth-resolved platform for blood-flow sensing. By estimating TOF-resolved speckle contrast through the g 1 -derived κ 2 route, TOF-iSCOS suppresses additive-noise bias while preserving sensitivity to deeper dynamic tissue layers. The present single-channel results bridge continuous-wave iSCOS, interferometric NIRS and time-domain diffuse correlation spectroscopy (TD-DCS), and motivate future multi-channel and cohort studies for scalable cortical hemodynamic monitoring.