Software-defined Radar for MRI Motion Correction: A versatile, vendor-independent Platform

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

To develop and evaluate a flexible, software-defined radar platform for contactless, vendor-independent motion detection and correction in MRI.

Methods

A continuous-wave (CW) Doppler radar was implemented using a software-defined radio and the open-source GNU Radio framework. The system was deployed inside a 1.5T MRI scanner and synchronized with MRI acquisitions. We evaluated the performance in a custom-developed internal motion phantom and in healthy volunteers to track respiration and bulk motion. The radar-derived signal was validated against cine MRI and used to demonstrate both retrospective and prospective motion management techniques in phantom and in healthy volunteers.

Results

The radar provided robust motion signals that correlated strongly with image-based ground truth signals in both phantom and volunteer experiments. Signal characteristics were found to be frequency-dependent, enabling optimization for different motion regimes. Retrospective correction of free-breathing abdominal data using the radar signal effectively suppressed respiratory artifacts, achieving image quality comparable to a self-gating approach. Prospective triggering successfully reduced motion artifacts in the phantom study. The system also reliably detected sporadic events such as swallowing during neck imaging.

Conclusion

Software-defined radar was demonstrated to be an effective platform for both prospective and retrospective motion correction. Its independence from the MRI system, ultra-wide band capabilities, and body-region versatility enable the adaptation of the technique for a wide range of imaging applications and protocols.