Evaluation of the Liberty16 Mobile Real Time PCR Device for Use With the SalivaDirect Assay for SARS-CoV-2 Testing
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
The COVID-19 pandemic has highlighted the need and benefits for all communities to be permitted timely access to on-demand screening for infectious respiratory diseases. This can be achieved with simplified testing approaches and affordable access to core resources. While RT-qPCR-based tests remain the gold standard for SARS-CoV-2 detection due to their high sensitivity, implementation of testing requires high upfront costs to obtain the necessary instrumentation. This is particularly restrictive in low-resource settings. The Ubiquitome Liberty16 system was developed as an inexpensive, portable, battery-operated single-channel RT-qPCR device with an associated iPhone app to simplify assay set-up and data reporting. When coupled with the SalivaDirect protocol for testing saliva samples for SARS-CoV-2, the Liberty16 device yielded a limit of detection (LOD) of 12 SARS-CoV-2 RNA copies/µL, comparable to the upper end of the LOD range for the standard SalivaDirect protocol when performed on larger RT-qPCR instruments. While further optimization may deliver even greater sensitivity and assay speed, findings from this study indicate that small portable devices such as the Liberty16 can deliver reliable results and provide the opportunity to further increase access to gold standard SARS-CoV-2 testing.
Article activity feed
-
-
SciScore for 10.1101/2021.08.08.21261746: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Ethics not detected. Sex as a biological variable not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Table 2: Resources
Software and Algorithms Sentences Resources Limit of Detection: A limit of detection (LOD) range finding study was conducted to compare a single-plex version of the SalivaDirect protocol3 run on the Liberty16 System as compared to the standard dualplex protocol run on the BioRad CFX96 Touch. SalivaDirectsuggested: NoneIn all other respects the assay was performed as described in the SalivaDirect protocol using the Thermo Fisher Proteinase K (A42363) and New England Biolabs Luna Universal Probe One-Step RT-qPCR (2x) kit (E3006). Thermo Fisher ProteinaseSciScore for 10.1101/2021.08.08.21261746: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Ethics not detected. Sex as a biological variable not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Table 2: Resources
Software and Algorithms Sentences Resources Limit of Detection: A limit of detection (LOD) range finding study was conducted to compare a single-plex version of the SalivaDirect protocol3 run on the Liberty16 System as compared to the standard dualplex protocol run on the BioRad CFX96 Touch. SalivaDirectsuggested: NoneIn all other respects the assay was performed as described in the SalivaDirect protocol using the Thermo Fisher Proteinase K (A42363) and New England Biolabs Luna Universal Probe One-Step RT-qPCR (2x) kit (E3006). Thermo Fisher Proteinasesuggested: (Resource Identification Portal, RRID:SCR_004098)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
Results from scite Reference Check: We found no unreliable references.
-