Sensitivity and specificity of rapid diagnostic tests: HAT Bioline Abbott® 2.0 and HAT Sero K-Set CORIS® 2.0 in the detection of Human African Trypanosomiasis

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Background Human African trypanosomiasis (HAT) due to Trypanosoma brucei gambiense has declined substantially in recent years, with many endemic areas now reporting very low prevalence levels as countries move toward elimination. In the Democratic Republic of the Congo (DRC), rapid diagnostic tests (RDTs) are increasingly used in surveillance and elimination programs; however, their performance in low-prevalence settings remains critical. This study evaluated and compared the diagnostic accuracy of two second-generation RDTs; HAT Bioline Abbott® 2.0 and HAT Sero K-Set CORIS® 2.0; using trypanolysis as the reference standard. Methods A retrospective diagnostic accuracy study was conducted on 447 archived plasma samples (227 confirmed HAT cases and 220 controls) from a previous study in the DRC. Each sample was tested with both RDTs and trypanolysis. Sensitivity, specificity, predictive values, likelihood ratios, Cohen’s kappa coefficient, and Youden’s index were calculated with 95% confidence intervals. Paired comparisons were performed using McNemar’s χ² test, with statistical significance set at p < 0.05. Results HAT Bioline Abbott® 2.0 showed a sensitivity of 85.2% (95% CI: 80.4–90.0) and specificity of 72.2% (95% CI: 66.4–77.9). The positive predictive value (PPV) was 73.1% (95% CI: 67.5–78.6) and the negative predictive value (NPV) was 84.7% (95% CI: 79.7–89.6). HAT Sero K-Set CORIS® 2.0 demonstrated a sensitivity of 90.0% (95% CI: 85.2–93.4) and a higher specificity of 88.2% (95% CI: 83.5–91.7). The PPV was 87.1% (95% CI: 82.0–90.9) and the NPV was 90.9% (95% CI: 86.4–93.9). HAT Sero K-Set CORIS® 2.0 showed stronger agreement with trypanolysis (κ = 0.78) and a higher Youden’s index (0.78) than Abbott® 2.0 (κ = 0.57; Y = 0.57). Conclusions HAT Sero K-Set CORIS® 2.0 showed superior overall diagnostic performance compared with HAT Bioline Abbott® 2.0. Although both RDTs demonstrated acceptable sensitivity, neither achieved the specificity threshold (>95%) recommended for surveillance and elimination verification in low-prevalence settings. These findings highlight the importance of optimizing specificity in serological screening tools as countries progress toward HAT elimination.

Article activity feed