Comparison of whole body bone mineral density measurement between dual-energy X-ray absorptiometry and novel foot-to-foot bioelectrical impedance analyzer

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Bone mineral density (BMD) is a crucial indicator of osteoporosis. Bioelectrical impedance analysis (BIA) introduces a new capability for assessing body composition, specifically BMD measurement. This study aimed to evaluate the accuracy of the novel BIA in conducting whole-body BMD tests in the general population of Taiwan. Altogether, 318 healthy adults in Taiwan (age, 37.67 ± 19.44 years; 145 male and 173 female patients) were included. Whole-body BMD was measured using foot-to-foot BIA-StarBIA201 (StarBIA Meditek Co. LTD, Taichung, Taiwan) and dual-energy X-ray absorptiometry (DXA) Lunar Prodigy (GE Medical Systems, Madison, WI, USA). Linear regression analysis, Pearson's correlation coefficient, Bland–Altman Plot, and paired t-test were used. Whole body BMD measured by BIA and DXA was 1.139 ± 0.124 g/cm 2 and 1.202 ± 0.168 g/cm 2 , respectively. The regression equation was y = 1.057x + 0.063. The Pearson correlation coefficient, mean difference, and limits of agreement were r  = 0.737, − 0.053 g/cm 2 , and − 0.290–0.165 g/cm 2 , respectively. Standing BIA was correlated with the DXA gold standard for estimating whole-body BMD in adults; however, their interchangeability remains limited. The convenient BIA method for measuring whole body BMD may be useful in the application of primary screening and future development of BMD assessment methods.