Evaluation of Different PCR Assay Formats for Sensitive and Specific Detection of SARS-CoV-2 RNA
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
Accurate identification of individuals infected with SARS-CoV-2 is crucial for efforts to control the ongoing COVID-19 pandemic. Polymerase chain reaction (PCR)-based assays are the gold standard for detecting viral RNA in patient samples and are used extensively in clinical settings. Most currently used quantitative PCR (RT-qPCRs) rely upon real-time detection of PCR product using specialized laboratory equipment. To enable the application of PCR in resource-poor or non-specialist laboratories, we have developed and evaluated a nested PCR method for SARS-CoV-2 RNA using simple agarose gel electrophoresis for product detection. Using clinical samples tested by conventional qPCR methods and RNA transcripts of defined RNA copy number, the nested PCR based on the RdRP gene demonstrated high sensitivity and specificity for SARS-CoV-2 RNA detection in clinical samples, but showed variable and transcript length-dependent sensitivity for RNA transcripts. Samples and transcripts were further evaluated in an additional N protein real-time quantitative PCR assay. As determined by 50% endpoint detection, the sensitivities of three RT-qPCRs and nested PCR methods varied substantially depending on the transcript target with no method approaching single copy detection. Overall, these findings highlight the need for assay validation and optimization and demonstrate the inability to precisely compare viral quantification from different PCR methodologies without calibration.
Article activity feed
-
-
SciScore for 10.1101/2020.06.24.168013: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources The recombinant plasmid was linearized using the Ncol restriction enzyme (ThermoFisher Scientific) and the MEGAscript® T7 Kit (ThermoFisher Scientific) was used for in vitro transcription of the fragment. MEGAscript®suggested: NoneResults from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the …SciScore for 10.1101/2020.06.24.168013: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources The recombinant plasmid was linearized using the Ncol restriction enzyme (ThermoFisher Scientific) and the MEGAscript® T7 Kit (ThermoFisher Scientific) was used for in vitro transcription of the fragment. MEGAscript®suggested: NoneResults from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- No funding statement was detected.
- No protocol registration statement was detected.
-