Effect of Heat inactivation on Real-Time Reverse Transcription PCR of the SARS-COV-2 Detection
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
Background: Real-time reverse transcription PCR (rRT-PCR) is commonly used to diagnose SARS-CoV-2 infection. Heat inactivation prior to nucleic acid isolation may allow safe testing, while the effects of heat inactivation on SARS-CoV-2 rRT-PCR detection result need to be determined. Methods: 14 positive nasopharyngeal swab specimens were inactivated at 56°C for 30min, 56°C for 60min, 60°C for 30min, 60°C for 75min, and 100°C for 10min, and another 2 positive nasopharyngeal swab specimens were also inactivated at 100°C for 10min, 100°C for 30min, 100°C for 60min, after which the samples were isolated and detected by rRT-PCR. Results: All 14 heat treated samples remained positive. The range of threshold cycle (Ct) values observed when detecting ORF1a/b was 27.228-34.011 in heat-treated samples, while 25.281-34.861 in unheated samples, and the range of threshold cycle (Ct) values observed at the time of detecting N was 25.777-33.351 in heat-treated samples, while 24.1615-35.433 in unheated samples, on basis of which it showed no statistical difference otherwise a good correlation of Ct values between the heat-inactivated samples and the untreated samples. However, the 2 samples inactivated at 100°C 30min, 100°C 60min turned into negative. Conclusions: Heat inactivation at 56°C for 30min, 56°C for 60min, 60°C for 30min, 60°C for 75min, and 100°C for 10min shall not affect the detection results of Real-Time Reverse Transcription PCR of the SARS-COV2. Furthermore, it is recommended to inactive nasopharyngeal swab specimens 10min at 100°C before RNA extraction in consideration of efficiency and reliable results. Key Words: SARS-CoV-2, Heat Inactivation, rRT-PCR, Comparison
Article activity feed
-
SciScore for 10.1101/2020.05.19.20101469: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources Graphpad Prism software (Version8.0) was used for statistical analysis. Graphpad Prismsuggested: (GraphPad Prism, RRID:SCR_002798)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: We detected the following sentences addressing limitations in the study:Our research has some limitations. First, only nasopharyngeal swab samples were …
SciScore for 10.1101/2020.05.19.20101469: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources Graphpad Prism software (Version8.0) was used for statistical analysis. Graphpad Prismsuggested: (GraphPad Prism, RRID:SCR_002798)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: We detected the following sentences addressing limitations in the study:Our research has some limitations. First, only nasopharyngeal swab samples were used for testing; which may make our results not applicable to those obtained using sputum, fecal specimen or blood samples. Second, we tested a relatively small number of samples under limit test condition, affecting the representative significance. Third, the thermostability of SARS-CoV-2 RNA in nasopharyngeal swab specimens at 100 °C for 10 min need to be determined. In the light of which, our findings need to be further verified by other large-scale researches and virus infectivity experiments using cell culture.
Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-
-