Utah-Stanford Ventilator (Vent4US): Developing a rapidly scalable ventilator for COVID-19 patients with ARDS
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
- Evaluated articles (preLights)
Abstract
We describe a minimum, rapidly scalable ventilator designed for COVID-19 patients with ARDS. Our design philosophy is not only to try to address potential ventilator shortages, but also to account for uncertainties in the supply chains of parts commonly used in traditional ventilators. To do so we employ a modular design approach and broadly explore taking advantage of parts from non-traditional supply chains. In our current prototype, we demonstrate volume control with assist control on a test lung and present a linear actuator-driven pinch valve-based implementation for both pressure control and volume control with decelerating inspiratory flow. We estimate the component cost of the system to be around $500. We publish our draft design documents and current implementation which is open and accessible in the hope that broadening the community globally will accelerate arriving at a solution and that peer review will improve the final design.
Article activity feed
-
SciScore for 10.1101/2020.04.18.20070367: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
No key resources detected.
Results from OddPub: Thank you for sharing your code.
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Th…
SciScore for 10.1101/2020.04.18.20070367: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
No key resources detected.
Results from OddPub: Thank you for sharing your code.
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-
Excerpt
Science and engineering meets disease: re-directing talents and ideas towards public health.
-