Bootstrapping surgeon labeling campaigns with x 0 -diffusion: label-efficient vessel segmentation of catheter-based angiograms
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Objective
Assess which architecture is the most label-efficient for vessel segmentation of catheter-based angiograms, and whether a diffusion model trained with extreme data scarcity could serve as a scaffold for expert annotation.
Materials and Methods
x 0 -diffusion, SAM, DeepLabV3, ResNet-UNet, ViT-DPT, and ResNet-ViT hybrid were compared at 16, 32, and 80 labeled images from the 500-image Dr-SAM dataset under 5×5-fold repeated cross-validation, using one-sided paired Wilcoxon signed-rank tests with Holm correction. A vascular surgeon’s annotation time, from scratch and when correcting x 0 -diffusion predictions, was also measured.
Results
At 16 training images, x 0 -diffusion achieved the highest Dice (0.882) and IoU (0.795) of the six models, outperforming every competitor in all 25 paired runs on Dice, by 0.047 over zero-shot SAM and 0.226 over DeepLabV3. The advantage narrowed as labels increased: at 32 images x 0 -diffusion was not significantly different from the two strongest convolutional models, and at 80 images those models scored marginally higher (0.923 and 0.920 vs. 0.910). Correcting 16-image x 0 -diffusion predictions on five test angiograms averaged 11.1 surgeon-minutes per image versus 20.6 from scratch.
Discussion
The advantage of x 0 -diffusion is label efficiency, not peak accuracy, with the strongest convolutional models needing twice the labels to match it. It rarely fails where zero-shot SAM fails, so at the start of an annotation campaign, correcting its predicted masks could cut surgeon annotation time by an estimated 37 to 45%.
Conclusion
With very few labels, x 0 -parameterized diffusion is the strongest vessel segmenter tested and a candidate scaffold for expert annotation.
Layman Summary
Segmenting vessels on an angiogram manually takes a vascular surgeon about twenty minutes per image. This study shows that an x 0 -diffusion model trained on only 16 angiograms draws outlines good enough to correct rather than redraw, which took the surgeon about half as long on five test images.