ACCURACY OF COMPRESSED SENSING ACCELERATED 3D-FLAIR IN THE DIAGNOSTIC WORK-UP OF MULTIPLE SCLEROSIS AT 1.5 T

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

BACKGROUND

3D-FLAIR is central to MR-based diagnosis and monitoring of MS, but long acquisition times remain a limitation.

OBJECTIVES

This single-center prospective non-inferiority study evaluated whether compressed sensing (CS) – accelerated 3D-FLAIR at 1.5T provides lesion detection comparable to optimized Generalized Autocalibrating Partially Parallel Acquisition (GRAPPA) – accelerated 3D-FLAIR. Optimized CS- and GRAPPA-accelerated 3D-FLAIR sequences were incorporated into MRI examinations in patients with known or suspected MS. The primary endpoint was manual lesion count by two blinded readers using a 10% non-inferiority margin. Secondary endpoints included image quality, SNR, CNR, and automated lesion count and volumetry.

RESULTS

Sixty-eight patients were included (mean age, 40.1±13.3 years; 61.8% women). CS-accelerated 3D-FLAIR reduced acquisition time by 31.8% compared with GRAPPA (4:00 vs 5:52 minutes). Non-inferiority for manual lesion count was demonstrated for both readers (reader A, p < .0001; reader B, p = .005). Interrater agreement was excellent for CS (ICC = 0.921; 95% CI, 0.837-0.957) and very good for GRAPPA (ICC = 0.851; 95% CI, 0.769-0.906). Subjective image quality was high for both techniques. GRAPPA showed higher objective signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR), whereas automated lesion counts and volumes were higher for CS.

CONCLUSION

CS-accelerated 3D-FLAIR provided non-inferior lesion detection with substantial scan-time reduction.