Imaging biological tissue at extended depth using swept-source visible-light OCT

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

We present a swept-source visible-light optical coherence tomography (SS-vis-OCT) system with twice the imaging depth of prior SS-vis-OCT and improved sensitivity for imaging biological samples, achieving shot-noise-limited operation. Because second-harmonic generation is limited by the booster optical amplifier’s output power, the visible-light SS power is capped at 450 µW. Increasing it further would require a more powerful, but costlier, amplifier. Instead, we reduce the receiver noise floor by selecting a balanced detector with NEP approaching the thermal (Johnson–Nyquist) limit. Real-time numerical k-linearization increases the effective number of linear-in-wavenumber samples per sweep, further improving SNR and extending the imaging range. Together, these improvements yield a sensitivity of 89 dB and an imaging depth of 9 mm in air or 6.5 mm in tissue, enabling full anterior-segment imaging of ex vivo porcine eyes.