Muscle and Lipid Suppression: MUSLIS MRI, a novel approach to musculoskeletal MRI

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

to introduce a novel magnetic resonance imaging (MRI) technique, MUSLIS MRI ( mus cle and li pid s uppression), that generates two co-registered 3D sets of images in a single acquisition: one set with lipid signal suppression (MUSLIS1) and a second set with muscle signal suppression (MUSLIS2).

Materials and Methods

Numerical simulations based on Bloch equations were performed to optimize the sequence parameters; the MP2RAGE sequence was employed for the implementation of MUSLIS imaging. MRI experiments were performed in vivo at 3 T. Quantitative measurements of signal contrast were carried out on MUSLIS1 and MUSLIS2. Various combination images and T1 parametric maps of the thighs were generated from the two co-registered 3D sets.

Results and Discussion

Optimal sequence parameters to achieve lipid suppression in one dataset (MUSLIS1) and muscle suppression in the other dataset (MUSLIS2) were determined by numerical simulations. MUSLIS imaging was validated in vivo at 3 T: MUSLIS1 images and MUSLIS2 images displayed excellent lipid and muscle suppression, respectively. The values of the signal contrast between subcutaneous lipids and muscle, as well as between bone marrow lipids and muscle reached values close to 0.8. Thus, MUSLIS would allow for tissue characterization of lesion/masses and direct visualization of the adipose tissue. Combination images, generated from these two intrinsically co-registered sets of images (MUSLIS1 and MUSLIS2), displayed different contrasts and improved signal characteristics, such as mitigation of signal inhomogeneities. The development of MUSLIS imaging for musculoskeletal imaging followed the concept introduced with the FLAWS, an MRI sequence designed for brain MR imaging to generate two 3D sets of images, one with white matter suppression and the other with cerebrospinal fluid suppression. Similarly to FLAWS, which provides T1 maps of the brain, MUSLIS imaging yields T1 parametric maps of the musculoskeletal system. Furthermore, the unique characteristics of the combination images make MUSLIS imaging a potentially attractive technique at field strengths higher than 3 T.

Conclusion

The proof-of-concept of a novel approach to musculoskeletal imaging, MUSLIS MRI, is proposed here and demonstrated by numerical simulations and MRI experiments performed in vivo at 3 T.