A public-data reproducibility benchmark of a 2.5-D U-Net with a slice-level topology term for type-B aortic dissection segmentation

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Background

Type-B aortic dissection requires separation of the true and false lumina in heterogeneous, anisotropic CTA. A topology term may reduce fragmentation, but its reproducibility must be tested under a fixed case split and independent native-grid evaluation.

Methods

We used the public Graz type-B aortic dissection CTA collection (40 cases; CC BY 4.0) and froze a 28/6/6 case split for training, development, and held-out testing. A compact 2.5-D U-Net trained with cross-entropy plus multiclass Dice was compared with the identical model augmented by a fixed-weight (0.5) differentiable slice-level soft clDice term. Two training seeds were run for each arm. All 40 cases passed the NRRD audit; predictions for six held-out cases were restored to their native grids and scored with case-level 3-D voxel-grid Dice, voxel-grid morphological clDice, 6-connected component counts, physical-volume error, and 2,000-resample case-level paired bootstrap intervals.

Results

With seed 20261007, topology-minus-baseline differences were −0.028 (95% CI −0.053 to −0.004) for true-lumen Dice, +0.032 (−0.006 to +0.096) for true-lumen voxel-grid clDice, +0.071 (+0.037 to +0.103) for false-lumen Dice, and +0.073 (+0.048 to +0.103) for false-lumen voxel-grid clDice. With seed 20261008, the corresponding differences were −0.040 (−0.065 to −0.018), +0.013 (−0.034 to +0.071), −0.099 (−0.187 to −0.004), and −0.058 (−0.124 to −0.001). Descriptive pooled means over 12 case rows per arm were true-lumen Dice 0.577 for topology versus 0.611 for baseline and false-lumen Dice 0.409 versus 0.422; these pooled rows are not independent cases.

Conclusions

Under this fixed six-case test split, the corrected slice-level topology term did not produce a consistent benefit. True-lumen Dice was lower in both seeds, while false-lumen effects reversed direction between seeds; topology-clDice changes were small and/or interval-crossing. The benchmark supports an auditable failure-analysis workflow but does not support a general topology benefit, clinical use, external generalization, or a panvascular decision system. The 3-D clDice quantity is an evaluation descriptor on the voxel grid; the training term is slice-level and does not enforce physical 3-D connectivity.

Current executed scope and claim boundary

This executed benchmark is restricted to the 40-case Graz type-B aortic dissection CTA collection with expert true- and false-lumen annotations. All 40 cases passed the audit; the frozen case split contains 28 training, 6 development, and 6 held-out test cases. Four runs (baseline/topology arms with seeds 20261007 and 20261008) produced native-grid predictions for all six held-out cases. The current estimand is descriptive case-level segmentation performance within this single public source. AVT, AortaSeg-60, ImageTBAD, and the ancillary multi-structure CCTA collection are documented as possible future external resources, but they are not part of the current endpoint, training set, or performance claim. This is a proof-of-concept segmentation benchmark, not a panvascular clinical system. The present study does not estimate multicentre generalisation, perform a TEVAR landing-zone assessment, evaluate device fit, or use follow-up outcomes. Any virtual-TEVAR geometry extension remains separately gated future work.