Validation of an electronic-nose sensor array in detecting C. difficile and differentiating toxin-positive from toxin-negative stool samples

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Objective

To evaluate the specimen-level performance of a multichannel electronic nose (e-nose) sensor array in classifying stool specimens into relevant Clostridioides difficile categories.

Design

Retrospective single-center validation study with reference to polymerase chain reaction (PCR) and C. difficile toxin testing.

Participants

A convenience sample of 31 discarded stool specimens with PCR and confirmatory toxin testing. Categorized into PCR-positive (+)/toxin+, PCR+/toxin-negative (–), and PCR–.

Methods

Specimens were sampled in controlled headspace by two e-nose sensor arrays in 10-minute sessions with 5-second sampling intervals. The primary endpoint was one-versus rest classification performance (sensitivity, specificity, positive predictive value, and accuracy) into three classes: PCR+/toxin+, PCR+/toxin–, and PCR–. Data were normalized and the region of interest was extracted for training and testing. Random forests were utilized for classification of normalized data with a test-train split at the specimen-level.

Results

There were 457 unique 10-minute sampling events (147 PCR+/toxin+, 183 PCR+/toxin–, 127 PCR–). Sensors 1 and 2 captured 20,331 and 23,538 unique 5-second exposures within the region of interest, respectively. Testing data classification accuracy was 0.79 [0.78, 0.81] and 0.84 [0.83, 0.85] for Sensors 1 and 2. Sensitivity for PCR+/toxin+ exposures was 0.82 [0.79, 0.84] and 0.94 [0.92, 0.95], and specificity was 0.91[0.90, 0.93] and 0.93 [0.92, 0.94] for Sensors 1 and 2.

Conclusion

An e-nose sensor array partially differentiated PCR-classified C. difficile specimen classes at the exposure level under headspace-controlled conditions. Performance was strongest for PCR+/toxin+ exposures. Findings support further prospective validation in the hospital room setting.