Virtual Noncontrast Imaging with Deep-Silicon Photon-Counting CT: Matched Phantom and Initial Human Evaluation

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Objectives

To compare virtual noncontrast (VNC) error between deep-silicon photon-counting CT (PCCT) and dual-energy energy-integrating detector CT (EID-CT) in a matched phantom and characterize tissue-specific error in humans.

Materials and methods

This prospective exploratory study included a matched phantom experiment and pilot human evaluation. Prototype projection-domain water-density maps served as the VNC surrogate. Twenty-four vials containing 0.1–20 mg I/mL in deionized water or simulated blood were scanned with PCCT and EID-CT. Three men underwent true noncontrast and contrast-enhanced PCCT; 80 matched measurements were analyzed. Paired t tests and mixed-effects models assessed system differences and iodine associations.

Results

Three participants (mean age, 67 ± 10 years; all men) were included. Mean VNC error was 29.1 HU lower with PCCT than EID-CT (95% CI, 15.5–42.7 HU; p < 0.001). Error was not associated with iodine concentration for PCCT (slope, 0.03 HU per mg I/mL; 95% CI, −0.92 to 0.98; p = 0.95) but increased for EID-CT (slope, 3.51 HU per mg I/mL; 95% CI, 2.56–4.46; p < 0.001). The slopes differed by 3.48 HU per mg I/mL ( p < 0.001). In humans, errors in five non-blood-pool tissues were within 5 HU. Error was not associated with iodine concentration among 61 non-blood-pool measurements (adjusted slope, −0.17 HU per mg I/mL; 95% CI, −1.22 to 0.88; p = 0.75).

Conclusion

PCCT provided VNC measurements with errors closer to zero and less dependent on iodine than EID-CT in the phantom. In humans, error was generally small in non-blood-pool tissues and was not associated with iodine concentration.

Key Points

Question

Can deep-silicon photon-counting CT provide virtual noncontrast measurements with errors close to zero and little iodine dependence in a matched phantom and human tissues?

Findings

Deep-silicon photon-counting CT showed phantom errors closer to zero and less dependent on iodine than dual-energy EID-CT; errors were generally small in non-blood-pool tissues.

Clinical Relevance Statement

Deep-silicon photon-counting CT provided virtual noncontrast measurements with little iodine dependence in non-blood-pool tissues, supporting evaluation in larger cohorts as an alternative to separate true noncontrast acquisitions.