Sensitivity of Pulmonary 4D Flow MRI Biomarkers to Controlled Spatial and Temporal Degradation in a CFD Benchmark

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

To quantify, in a computational fluid dynamics (CFD) benchmark, how spatial degradation, temporal acquisition and noise change the error of plane-based pulmonary-artery flow endpoints.

Methods

Twenty-two historical, non-converged CFD simulations (11 geometry families) were voxelised at 2 mm with 80 frames per cycle. Simulations with numerical divergence were excluded, leaving 13 (7 families). Fields were degraded to 4 and 8mm and with noise, plane waveforms were sampled at 20 phases with one-interval averaging, and errors were computed against the noise-free reference waveform on 66 geometric planes. Family-mean paired differences are reported with t-based 95% confidence intervals under an analysis plan fixed in advance.

Results

Spatial degradation from 2 to 4mm increased full-waveform nRMSE by 0.0032 [0.0021, 0.0043], in 7 of 7 families; cycle-mean flow bias accounted for 23% of the squared error, against 60% at 8 mm. With SNR scaled to voxel volume, the increase was not resolved (0.0002 [−0.0012, 0.0015]). Twenty-phase acquisition increased PI error (0.0223 [0.0214, 0.0233]) and diastolic-trough timing error (0.63 [0.52, 0.74] frames) more than spatial degradation, and its point estimates for these endpoints were larger with every recovery method; the nRMSE ordering and peak timing depended on the recovery method. The implementation of the downsampling operator mattered: an alternative, corner-aligned implementation of the same 4mm degradation gave a larger spatial effect (nRMSE difference 0.0010 [0.0002, 0.0017]).

Conclusion

In this benchmark, a 2 to 4mm degradation produced a small, consistent, mostly shape-related nRMSE increase that depended on the noise model, whereas 20-phase acquisition changed PI error and trough timing more. The results describe benchmark behaviour, not clinical acquisition requirements.