Quantitative Fetal Hemoglobin Measurement: Gazelle™ Point-of-Care Electrophoresis versus Standard HPLC Methodology
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Quantitative measurement of fetal hemoglobin (HbF) using high-performance liquid chromatography (HPLC) is standard but expensive and widely unavailable in low-income settings. The newer Gazelle™ method, which employs single-use alkaline electrophoresis cartridges, provides a portable and relatively inexpensive potential alternative. In the current study, we tested 3055 venous blood samples collected from 1780 children with sickle cell anemia who were enrolled into the H-PRIME trial by both Gazelle and the Bio-Rad D-10 HPLC instrument, before and during hydroxyurea treatment. Data reinterpretation was required to quantify accurately high HbF values derived using the D-10. Mean (standard deviation) baseline HbF values by HPLC and Gazelle (N=2042) were 11.4% (SD 5.9%) versus 16.3% (SD 7.5%), respectively, while mean on-treatment HbF values (N=1013) were 18.9% (SD 9.0%) versus 25.4% (SD 11.5%), respectively. While overall agreement was good (Lin’s concordance 0.76, 95% CI 0.75-0.78), HbF values derived using Gazelle were modestly higher than HPLC values (Bland-Altman agreement mean +5.4%, 95%CI −2.4% to +13.2%), with the difference between measures increasing at higher HbF percentages. With strong concordance and more easily interpretable HbF values, Gazelle represents an attractive and inexpensive alternative to HPLC for HbF quantification in the low-income settings where most patients with sickle cell anemia live.
Key Points
Point-of-care Gazelle electrophoresis compares favorably to standard HPLC hemoglobin separation to quantify fetal hemoglobin.
Gazelle HbF quantitation is accurate, inexpensive, and suitable for use in low-income settings including sub-Saharan Africa.