K-space Optimized Spatial Temporal Architecture for Low Latency in Interventional Cardiac MRI
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Purpose
To develop and evaluate a recurrent neural network that operates directly on undersampled golden-angle radial k-space, enabling low-latency cardiac cine reconstruction for interventional cardiac MRI (iCMR).
Methods
KOSTAL (K-space Optimized Spatial Temporal Architecture for Low latency) combines a per-coil k-space convolutional gated recurrent unit (convGRU), which learns a mask-conditioned view-sharing rule over the accumulated k-space, with an image-space recurrent module. It was trained and tested on the open-access multi-coil OCMR dataset using 10 simulated golden-angle spokes per frame against a 1000-spoke reference. Ablations varied recurrent module architecture and placement; temporal behavior was assessed using a imaging-plane transition test. Generalization was evaluated without retraining in 23 pediatric patients prospectively scanned during free-breathing using a radial golden-angle bSSFP sequence with binned XD-GRASP and clinical breath-held cine serving as references.
Results
At 10 spokes per frame, KOSTAL reached a median SSIM of 0.907 (gridding: 0.622) in under 34 ms of processing time per frame. Image quality (IQ) matched gridded results formed from 7-11 times more spokes. Recurrence in k-space rather than image space accounted for most of this gain (0.891 versus ≤0.650). IQ recovered from an abrupt plane transition in 75 ms, versus 303 ms for 200-spoke gridding. KOSTAL had the best agreement with the binned reference and the highest long-axis CNR. KOSTAL accurately tracked left-ventricular blood-pool volume (end-diastolic volume bias-0.7 mL) with a shorter lag (71 ms) than iterative reconstruction (239-287 ms).
Conclusion
Learned, mask-conditioned recurrence in k-space yields high image quality at a latency compatible with interventional guidance.