Accelerated 2-Minute Multi-Echo UTE MRI for Simultaneous CT-Like Bone Imaging and Quantitative Short-T2* Mapping

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

To determine the feasibility of a 2-minute multi-echo UTE (mecho-UTE) for CT-like bone-weighted contrast and T2* quantification of tissues with short T2/T2*.

Methods

Mecho-UTE data acquired from four patients and five healthy subjects were used to assess image quality of the CT-like contrast. All data were reconstructed using conventional gridding (GRID+CONV) and compared with those reconstructed using conjugate gradient SENSE combined with deep learning-based denoising (CG+DLR). Image resolution and sharpness of the CT-like images were assessed using the full width at half maximum (FWHM) and relative edge sharpness (RESH), respectively. Calimetrix UTE-T2* phantom was used to assess the accuracy of T2* quantification of the mecho-UTE sequence.

Results

Two-minute mecho-UTE with CG+DLR has similar accuracy (0.37 ± 0.27 vs. 0.67 ± 0.54 ms, p=0.20) and better precision (0.28 ± 0.16 vs. 1.23 ± 0.29 ms, p<0.001) compared to the 5-minute mecho-UTE with GRID+CONV. The 2-minute mecho-UTE with CG+DLR has higher resolution and sharpness compared to the 5-minute scan with GRID+CONV.

Conclusion

It is feasible to achieve simultaneous CT-like contrast and T2* quantification of short-T2 tissues in two minutes. When appropriately used, it may simplify logistics, reduce costs, and eliminate radiation exposure risks.