BioDeformUNet: A Deep Learning Model for Biomechanically Informed Liver Image Registration

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

To build a 3D U-Net model, BioDeformUNet, to predict the liver’s deformation vector field (DVF) in near real-time, for efficient intra-procedural evaluation of the minimal ablative margin (MAM).

Materials and Methods

This retrospective study included 170 contrast-enhanced computed tomography (CECT) image pairs from 157 patients who underwent liver ablation treatment between 2020-2024. Each data instance included one pre-ablation CECT (pre-CECT) and one post-ablation CECT (post-CECT). BioDeformUNet was trained under the guidance of DVFs generated by a biomechanical model-based deformable image registration (DIR) algorithm using a loss function that focused on large liver deformations. Data were split patient-wise into training (92-93 patients), validation (23-24 patients), and testing sets (42 patients). We compared our method’s performance with two deep learning-based DIR methods: VoxelMorph and VFA. Evaluation metrics included: target registration error (TRE), Dice similarity coefficient (DSC), Minimum Ablation Margin (MAM), and inference time. For BioDeformUNet, we additionally evaluated the accuracy of the deformed tumor center-of-mass mapping by comparing the predicted tumor center location with that generated by Morfeus. A mapping error less than 3.0 mm (corresponding to the voxel size) was considered accurate. We used the Wilcoxon signed-rank test to assess the significancy of each test result. Our code is available at https://github.com/XinyueZhang831/BioDeformUNET .

Results

The TRE of BioDeformUNet was not significantly different from Morfeus (3.31±1.17 BioDeformUNet; 3.23±1.06 Morfeus; p-value=0.41). The BioDeformUNet DVF magnitude was within 3.0 mm of Morfeus DVF for an average of 91.9 ± 17.6% (mean ± SD) of the voxels. Tumor mapping errors greater than 3.0 mm occurred in only 8 cases. The inference time of BioDeformUNet was 0.6s per image pair, 0.2s for VoxelMorph, 0.3s for VFA, and 20.2s for Morfeus.

Conclusion

BioDeformUNet achieved a similar performance to the biomechanical model-based algorithm but required fewer computational operations, resulting in a 34× speedup in DVF computation.

Article activity feed