Development of a Matrix-Matched Calibration Curve for Multi-Site Quantification of Neu5Gc-Bearing N-Glycans

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

N -glycolylneuraminic acid (Neu5Gc) has been repeatedly associated with human cancer, but reliable detection has remained elusive, generating controversy regarding its presence in human samples. To address this, matrix-matched calibration curves, which have been pioneered in proteomics and metabolomics for assessing changes in complex mixtures, were measured of released N -glycans at four orders of magnitude dynamic range in defined mixtures, systematically benchmarking Neu5Gc-containing N -glycan detection across multiple LC-MS platforms and sites. Orthogonally, the gold-standard analytical method, consisting of fluorescence detection of labelled monosaccharides separated by LC, was applied to the same samples, yielding absolute concentrations of Neu5Gc.

LC-MS demonstrated an extended detection range of three or more orders of magnitude while retaining intact N -glycan measurement, improving assay specificity and enabling detection of the variety of Neu5Gc-bearing N -glycans. By combining orthogonal dimensions of evidence, including chromatographic separation, isotopic distribution matching, and composition-confirming MS/MS, LC-MS confidently resolved Neu5Gc signals from noise, even at low abundance. In comparison, DMB-LC-FLR was limited to two orders of magnitude dynamic range, insufficient for detection of Neu5Gc in commercially available pooled human sera. These findings strongly support that DMB-LC-FLR assay specificity and sensitivity are insufficient for Neu5Gc detection in human samples due to noise overwhelming the Neu5Gc signal. By establishing a reusable benchmarking framework for future glycomic studies, we aim to use LC-MS to improve the measurement of Neu5Gc in clinical samples.

Article activity feed