Intravital mid-infrared biosensing by normalized spatial probing of self-referenced optothermal signals

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

We present a compact pump-and-probe mid-infrared Optothermal Spectrometer (OTHES) equipped with Spatial Probing and Autocorrection (SPAC) optimized for robust intravital application in humans. SPAC-OTHES facilitates alignment stability and spectral comparability across different measurement sessions involving different skin types. Contrary to state-of-the-art, SPAC-OTHES uses camera-based beam detection and an auto-calibration mechanism that enables ca. 73% better spectral reproducibility in intravital measurements in human volunteers than non-calibrated readouts. Moreover, SPAC-OTHES has the potential to lower the glucose quantification error, as demonstrated here in artificial skin phantoms, where an improvement of 52% compared to conventional diode-based detection was observed. The compactness of OTHES, combined with reliable SPAC-readout, has the potential to accelerate commercialization and broad application of biosensors based on mid-infrared spectroscopy.

Article activity feed