An AI-Powered Smartphone Application for Universal and Standardized Reading and Interpretation of Lateral Flow Assays
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Introduction
Lateral flow assays (LFAs) are indispensable rapid diagnostic tools in healthcare, enabling point-of-care diagnosis critical for patient management and support disease burden assessment and surveillance when results are properly recorded. However, misinterpretation errors and unreported cases remain a concern. A quality-assured, affordable Ai-powered tool, supporting the decision-making during result interpretation could promote proper disease monitoring and epidemiological surveillance. Here, we describe the performance of a universal AI model to digitize and interpret results from multiple LFA types through a smartphone application, a step that could ultimately enable standardized and digitally reportable test outcomes.
Methods
The AI algorithm was evaluated in 17 LFA types, including both 2-band and 3-band tests for different diseases and manufacturers. The model was trained on a dataset of 22,576 images captured under diverse lighting conditions with different smartphone models and using a custom mobile application, TiraSpot (Spotlab, Madrid, Spain). To assess generalizability, a leave-one-out cross-validation was applied, wherein each LFA type was iteratively excluded from training and used for testing. Model performance was evaluated using bootstrapping on the inference dataset.
Results
In the assessment of the model’s ability to generalize to new LFA types not previously analyzed (not included during development), the model achieved an overall AUC of 94.3% for second band detection. This overall performance was enhanced to 99.3% (Sensitivity=98,6%; Specificity=98%) after training with 50 images of each LFA type, highlighting the benefit of additional data for specific LFA types. For the third band detection, where less training data was available, the system achieved an overall AUC of 83.9% for unseen LFAs, improving to 94.2% (Sensitivity=92.9%; Specificity=87,9%) after training with 50 images of each LFA type.
Conclusion
This system demonstrates the feasibility of an AI-powered universal digital reader for interpreting LFA results from diverse test types using smartphone-captured images. Its compatibility with standard smartphones makes it a universal tool, enabling reliable LFA interpretation across devices and settings. By standardizing test interpretation and digitizing results, this tool could support decision making in result interpretation, enhancing epidemiological surveillance, particularly in resource-limited settings. Its adaptability across various infections highlights its potential to improve diagnostic consistency and support disease management in diverse healthcare settings.