Method and System for Heart Rate Estimation Using Linear Prediction Filtering
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Cardiovascular diseases represent one of the major problems faced by modern society. In addition to reducing people’s quality of life, bringing high costs to the health system, and causing losses in economic productivity, they are the leading cause of death in the world. Early diagnosis and treatment are the best actions to minimize the damage and costs caused by these diseases. For this, developing techniques and technologies that have higher accuracy in the analysis of electrocardiogram (ECG) signals is necessary. Early diagnosis benefits from relevant ECG interpretation. Then, it can contribute to reducing healthcare costs by replacing interventionist responses with preventive actions. This work presents a method and system for heart rate estimation using Linear Prediction Coefficients (LPCs) centered on an ESP32 microprocessor module and an AD8232 ECG signal conditioning module. The proposal was validated with a Tektronix AFG1022 function generator that produces ECG signals and obtained measurements with accuracy above 98.87%, showing performance similar to studies presented in the literature. Also, the LPC algorithm showed good performance in rejecting low-frequency noise caused by some common artifacts, such as body movement and electrode displacement.