Analytical Performances of the COVISTIXTM Antigen Rapid Test for SARS-CoV-2 Detection in an Unselected Population (All-Comers)
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
The performance and validity of the COVISTIXTM rapid antigen test for the detection of SARS-CoV-2 were evaluated in an unselected population. Additionally, we assessed the influence of the Omicron SARS-CoV-2 variant in the performance of this antigen rapid test. Swab samples were collected at two point-of-care facilities in Mexico City from individuals that were probable COVID-19 cases, as they were either symptomatic or asymptomatic persons at risk of infection due to close contact with SARS-CoV-2 positive cases. Detection of the Omicron SARS-CoV-2 variant was performed in 91 positive cases by Illumina sequencing. Specificity and sensitivity of the COVISTIXTM rapid antigen test was 96% (CI 95% 94–98) and 81% (CI 95% 76–85), respectively. The accuracy parameters were not affected in samples collected after 7 days of symptom onset, and it was possible to detect almost 65% of samples with a Ct-value between 30 and 34. The COVISTIXTM antigen rapid test is highly sensitive (93%; CI 95% 88–98) and specific (98%; CI 95% 97–99) for detecting Omicron SARS-CoV-2 variant carriers. The COVISTIXTM rapid antigen test is adequate for examining asymptomatic and symptomatic individuals, including those who have passed the peak of viral shedding, as well as carriers of the highly prevalent Omicron SARS-CoV-2 variant.
Article activity feed
-
-
SciScore for 10.1101/2021.09.10.21263410: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Ethics IRB: The study was approved by the ethics and research committees of Instituto Nacional de Medicina Genómica (CEI/1479/20 and CEI 2020/21). Sex as a biological variable not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Table 2: Resources
Software and Algorithms Sentences Resources We ran all tests with Thermo Fisher’s ABI QuantStudio 5 or QuantStudio 7 real-time thermal cyclers. Thermo Fisher’ssuggested: NoneResults from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques …SciScore for 10.1101/2021.09.10.21263410: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Ethics IRB: The study was approved by the ethics and research committees of Instituto Nacional de Medicina Genómica (CEI/1479/20 and CEI 2020/21). Sex as a biological variable not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Table 2: Resources
Software and Algorithms Sentences Resources We ran all tests with Thermo Fisher’s ABI QuantStudio 5 or QuantStudio 7 real-time thermal cyclers. Thermo Fisher’ssuggested: NoneResults from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
Results from scite Reference Check: We found no unreliable references.
-