Performance of Qualitative and Quantitative Antigen Tests for SARS-CoV-2 Using Saliva
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
The rapid detection of SARS-CoV-2 is critical for the prevention of disease outbreaks. Antigen tests such as immunochromatographic assay (ICA) and chemiluminescent enzyme immunoassay (CLEIA) can yield results more quickly than PCR. We evaluated the performance of ICA and CLEIA using 34 frozen PCR-positive (17 saliva samples and 17 nasopharyngeal swabs [NPS]) and 309 PCR-negative samples. ICA detected SARS-CoV-2 in only 14 (41%) samples, with positivity rates of 24% in saliva and 59% in NPS. Notably, ICA detected SARS-CoV-2 in 5 of 6 samples collected within 4 days after symptom onset. CLEIA detected SARS-CoV-2 in 31 (91%) samples, with a positivity of 82% in saliva and 100% in NPS. These results suggest that the use of ICA should be limited to an earlier time after symptom onset and CLEIA is more sensitive and can be used in situations where quick results are required.
Article activity feed
-
-
SciScore for 10.1101/2020.11.06.20227363: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement IRB: This study was approved by the Institutional Ethics Board (Hokkaido University Hospital Division of Clinical Research Administration Number: 020-0116), and informed consent was obtained from all patients.
Consent: This study was approved by the Institutional Ethics Board (Hokkaido University Hospital Division of Clinical Research Administration Number: 020-0116), and informed consent was obtained from all patients.Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Antibodies Sentences Resources In brief, the treatment solution and the specimen were consecutively aspirated using … SciScore for 10.1101/2020.11.06.20227363: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement IRB: This study was approved by the Institutional Ethics Board (Hokkaido University Hospital Division of Clinical Research Administration Number: 020-0116), and informed consent was obtained from all patients.
Consent: This study was approved by the Institutional Ethics Board (Hokkaido University Hospital Division of Clinical Research Administration Number: 020-0116), and informed consent was obtained from all patients.Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Antibodies Sentences Resources In brief, the treatment solution and the specimen were consecutively aspirated using a single-use tip and was dispensed into suspension of magnetic beads coated with the monoclonal antibody to SARS-CoV-2 N-Ag. SARS-CoV-2suggested: NoneN-Agsuggested: (Bethyl Cat# A303-897A, RRID:AB_2620247)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-