Interdevice Agreement of Keratometry, Astigmatism Vectors, and Ocular Biometry in Cataract Candidates: SS-OCT (Argos) vs. OLCI (Aladdin) vs. Scheimpflug–Placido (Sirius)

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Background and Objectives: Accurate anterior segment measurements are central to intraocular lens (IOL) power calculation and toric planning, yet different optical platforms may yield non-interchangeable values. This study compared keratometry, astigmatism metrics, and ocular biometry obtained with a swept-source OCT biometer (Argos), an optical low-coherence interferometry biometer (Aladdin), and a combined Scheimpflug–Placido topographer (Schwind Sirius). Methods: This is a retrospective observational study (January 2022–June 2024) including eyes undergoing uncomplicated cataract surgery. All eyes were measured in a single session by one examiner. Outcomes included K1, K2, cylinder, astigmatism axis (degrees; device-reported corneal cylinder axis, labeled “Powerful Angle” in the Sirius export), vector components (J0 and J45), and—where available—lens thickness (LT), axial length (AL), anterior chamber depth (ACD), white-to-white (WTW) distance, and central corneal thickness (CCT). Friedman tests assessed 3-device differences, and pairwise comparisons were evaluated using Wilcoxon signed-rank tests (paired data). Results: A total of 170 eyes (102 patients) were analyzed (mean age: 69.12 ± 10.26 years). Significant inter-device differences were detected for K1 (Argos: 43.45 ± 1.64 D; Aladdin: 43.41 ± 1.70 D; overall: p < 0.001; Argos vs. Aladdin: p = 0.019), K2 (Argos: 44.45 ± 1.67 D; Aladdin: 44.34 ± 1.71 D; overall and pairwise: p < 0.001), and cylinder (Argos: −0.83 ± 0.74 D, Aladdin: −0.77 ± 0.76 D; Sirius: −0.68 ± 0.75 D; overall: p < 0.001). “Powerful Angle” differed across devices (p = 0.003) but not between Argos and Aladdin (p = 0.512). J0 (p = 0.277) and J45 (p = 0.084) did not differ significantly. Argos reported higher ACD (3.19 ± 0.42 vs. 3.13 ± 0.41 mm, p < 0.001) and WTW (11.95 ± 0.42 vs. 11.65 ± 0.39 mm, p < 0.001) values than Aladdin. CCT was similar between Aladdin and Sirius (540.27 ± 33.44 vs. 540.47 ± 33.78 µm, p = 0.169). Conclusions: Several keratometric and biometric parameters differed significantly by device, indicating limited interchangeability—particularly relevant for toric and premium IOL planning—while vector astigmatism components and CCT showed better agreement.

Article activity feed