Whole-body tomographic fluorescence lifetime imaging of PD-L1 expression
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Noninvasive quantification of immune markers such as programmed death ligand-1 (PD-L1) remains a major challenge because conventional molecular imaging methods cannot readily distinguish target-bound from nonspecifically retained probes within the tumor microenvironment. Here, we establish a framework for noninvasive quantification of PD-L1 expression in deep-seated hepatocellular carcinoma (HCC) using asymptotic time-domain (ATD) fluorescence lifetime tomography. In vivo time-domain fluorescence imaging was performed in mice bearing orthotopic HCC tumors following administration of a PD-L1-targeted near-infrared fluorescent probe (αPDL1-800). Multi-exponential analysis of time-domain fluorescence data was used to derive four amplitude-based metrics of PD-L1 expression. Quantitative in vivo measurements obtained from ATD tomography were validated against ex vivo PD-L1 expression measured by Western blotting. Among the evaluated metrics, a normalized parameter that accounts for inter-lifetime crosstalk demonstrated the strongest correlation with ex vivo PD-L1 expression (r 2 = 0.77). These findings establish a foundation for a noninvasive, nonionizing imaging approach to quantify and monitor receptor expression and support a future path for the longitudinal assessment of immune biomarkers in both preclinical studies and clinical settings.