Mechanical Effects of Taper Angles in Implant–Abutment Connection: A Finite Element Study
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Purpose To analyze how the taper angle (defined here as the half‑angle per side) influences the mechanics of the implant–abutment connection using finite element analysis. Methods Three‑dimensional finite element models (implant body, abutment, and screw) with taper angles of 8°, 15°, 18°, and 22° were established in ABAQUS/CAE. All components were modeled as linearly elastic Ti‑6Al‑4V (E = 110 GPa, ν = 0.35). Frictional contact (µ = 0.3) was assigned at the taper and screw interfaces. A bolt load of 605 N (equivalent to 35 Ncm) was applied and then released to assess the press‑fit retention. Under the maintained preload, 100-N vertical and horizontal loads (unidirectional and bidirectional) were applied for five cycles. The primary outcomes are the abutment displacement, implant von Mises stress, bolt load change, and microgap size. Results Smaller taper angles (8°, 15°) retained press‑fit after preload release, whereas larger angles (18°, 22°) lost press‑fit contact. As the taper angle decreased, the abutment sinking, implant stresses, and bolt load loss increased. The effect of the loading direction follows the order: vertical < horizontal (unidirectional) < horizontal (bidirectional). Microgaps decreased with smaller taper angles and cycling. Conclusion Taper angle differences affect press-fit, abutment displacement, screw loosening, stress distribution, and microgap formation.