Diagnostic Accuracy and Practicability of the VISITECT® Advanced HIV Disease CD4 Test Compared to the Partec® CyFlow Counter II for CD4 Enumeration in Nigeria’s HIV Program
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
We evaluated the diagnostic performance and practicability of the VISITECT® CD4 Lateral Flow Assay for identifying advanced HIV disease (AHD) across 334 facilities in Nigeria. In two phases, 468 participants were tested with VISITECT® CD4 and Partec® CyFlow. VISITECT® CD4 achieved a sensitivity of 100.0% and specificity of 89.8% for detecting CD4 ≤ 200 cells/mm³. The positive predictive value (PPV) was 91.3% (95% CI: 87.4–94.1), and the negative predictive value (NPV) was 100% (95% Cl: 98.1–100.0). Overall diagnostic accuracy was 95.1% (95% CI: 92.8–96.8). We recommend its use to enhance timely AHD detection and improve linkage to prophylactic interventions in resource-limited settings.