Ureteral stent sleeve for early detection of hydronephrosis

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Hydronephrosis, a serious complication of ureteral stents, can lead to deterioration of renal function due to prolonged increased intrarenal pressure. Currently the only way to diagnose hydronephrosis is radiographic imaging and a continuous, non-invasive method is not available. This study introduces the ureteral stent sleeve, UroSleeve, a modular wireless pressure monitoring system that is designed to integrate seamlessly with standard ureteral stents without altering their existing designs or manufacturing processes. The UroSleeve incorporates a flexible printed-circuit-board-based spiral antenna and a surface-micromachined capacitive pressure sensor, forming an inductor-capacitor tank circuit capable of wireless telemetry through near-field inductive coupling. The device was evaluated in an ex vivo porcine kidney model whose internal pressure was externally controlled to mimic a hydronephrosis condition, and kidney pressure changes were successfully monitored by detecting shifts in the resonant frequency of the device. Key results demonstrated a high sensitivity with -5.3 ± 0.74 kHz/mmHg with a baseline resonant frequency of 15.234 MHz at 8.5 mmHg. The frequency response showed a strong correlation with kidney pressure, and the device maintained consistent performance in a relevant biological environment. The results indicate that UroSleeve offers a practical and versatile solution for wireless intrarenal pressure monitoring amendable for clinical translation.