Validation of Guiding Needle Placement Using Registered Ultrasound Imaging for Gynecologic Brachytherapy in a Simulated Pelvic Phantom
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Our previous research demonstrated that, under ideal conditions, high-precision image registration between real-time ultrasound (US) images and preoperative CT/MR images could be achieved using real-time US guidance for needle insertion. In this study, we enhanced a constructed phantom to better simulate the structure of human abdominal tissue, aiming to verify the accuracy of surgical needle placement guided by real-time US images registered with preoperative CT/MR images. Specifically, the pelvic phantom was improved using animal abdominal tissue to mimic the properties of human abdominal tissue. In the improved phantom experiments, the target registration error (TRE) for the real-time needle trajectory of four implanted needles in the registered US-CT images was 0.81 ± 0.11 mm (mean ± SD), with an average trajectory angle error of 0.83 ± 0.063 degrees (mean ± SD). For the registered US-MR images, the average registration error was 0.76 ± 0.014 mm (mean ± SD), and the average trajectory angle error was 0.825 ± 0.018 degrees (mean ± SD). These results demonstrate that, even in simulated human tissue, rigid registration between real-time US images and preoperative CT/MR images maintained high accuracy, supporting the feasibility of real-time US-guided needle insertion.