Evaluating the Accuracy of HIV-1 Viral Load Testing Using near point of care Xpert HIV- 1 System at Solwezi General Hospital, Zambia.

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Objective HIV-1 Viral Load Testing plays a crucial role in the management of HIV/AIDS patients by quantifying the presence of HIV-1 RNA in the blood, which directly relates to viral replication and disease progression. Monitoring viral load enables healthcare professionals to evaluate the effectiveness of Anti-Retroviral Therapy, adapt treatment strategies, and assess the overall health status of individuals with HIV. This was a cross-sectional study to evaluate analytical performance (accuracy, precision and linearity of the near point of care Xpert HIV-1 viral load. Results The analytical performance of the GeneXpert assay to the conventional Hologic Panther system was strong. Specifically, the GeneXpert demonstrated a sensitivity, specificity, and accuracy of 87.6%, 100%, and 98.5%, respectively, when compared to the Hologic Panther. A strong correlation between the two assays was evident (r = 0.97, p < 0.0001). Additionally, precision, as indicated by the coefficient of variation (CV), for high viral load was 1.38% for GeneXpert while for low viral load, it was 2.59%. In the linearity analysis, all data points remained within acceptable ranges, yielding correlation coefficients ≥ 0.99 for both assays. Conclusion The Xpert HIV-1 viral load assay demonstrates high efficacy in accurately identifying and quantifying HIV-1 viral loads, making it a valuable tool for HIV diagnosis and monitoring. This reliability makes it particularly suitable for critical populations requiring timely and accurate viral load monitoring.

Article activity feed