Reliability of an all-in-one wearable sensor for continuous vital signs monitoring in high-risk patients: the NIGHTINGALE clinical validation study
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Purpose Continuous vital signs monitoring with wearable systems may improve early recognition of patient deterioration on hospital wards. The objective of this study was to determine whether the wearable Checkpoint Cardio’s CPC12S, can accurately measure heart rate (HR), respiratory rate (RR), oxygen saturation (SpO2), blood pressure (BP) and temperature continuously. Methods In an observational multicenter method comparison study of 70 high-risk surgical patients admitted to high-dependency wards; HR, RR, SpO2, BP and temperature were simultaneously measured with the CPC12S system and with ICU-grade monitoring systems in four European hospitals. Outcome measures were bias and 95% limits of agreement (LoA). Clinical accuracy was assessed with Clarke Error Grid analyses for HR and RR. Results A total of 3212 hours of vital signs data (on average 26h per patient) were analyzed. For HR, bias (95% LoA) of the pooled analysis was 0.0 (-3.5 to 3.4), for RR 1.5 (-3.7 to 7.5) and for SpO2 0.4 (-3.1 to 4.0). The CPC12S system overestimated BP, with a bias of 8.9 and wide LoA (-23.3 to 41.2). Temperature was underestimated with a bias of -0.6 and LoA of -1.7 to 0.6. Clarke Error Grid analyses showed that adequate treatment decisions regarding changes in HR and RR would have been taken in 99.2% and 92.0% of cases respectively. Conclusion The CPC12S system showed high accuracy for measurements of HR. The accuracy of RR, SpO2 were slightly overestimated and temperature slightly underestimated, but were reasonably accurate and within the predefined clinically acceptable limits. The accuracy of BP was unacceptably low.