PhotoFiTT: A Quantitative Framework for Assessing Phototoxicity in Live-Cell Microscopy Experiments

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Phototoxicity in live-cell fluorescence microscopy can compromise experimental outcomes, yet quantitative methods to assess its impact remain limited. Here we present PhotoFiTT (Phototoxicity Fitness Time Trial), an integrated framework combining a standardised experimental protocol with advanced image analysis to quantify light-induced cellular stress in label-free settings. PhotoFiTT leverages machine learning and cell cycle dynamics to analyse mitotic timing, cell size changes, and overall cellular activity in response to controlled light exposure. Using adherent mammalian cells, we demonstrate PhotoFiTT's ability to detect wavelength- and dose-dependent effects, showcasing that near-UV light induces significant mitotic delays at doses as low as 0.6 J/cm 2 , while longer wavelengths require higher doses for comparable effects. PhotoFiTT enables researchers to establish quantitative benchmarks for acceptable levels of photodamage, facilitating the optimisation of imaging protocols that balance image quality with sample health.

Article activity feed