Large-field high-resolution X-ray AKB microscope for measuring hydrodynamic instabilities at the SG-Ⅲ prototype Laser Facility
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
X-ray imaging with a large field of view (FOV) and high resolution is extremely important for Rayleigh–Taylor instability measurement with a small amplitude and high spatial frequency in laser inertial confinement fusion. We developed an advanced Kirkpatrick–Baez (AKB) microscope based on the quadratic-aberration theory to realize a large FOV and high resolution. This microscope was assembled and tested in a laboratory, and it was then successfully applied for imaging the hydrodynamic instability of a perturbation target in implosion experiments at the Shenguang-III prototype laser facility. Imaging results demonstrate that the AKB microscope can achieve an optimal resolution of ~ 0.53μm and ~ 0.40μm and a spatial resolution of <1.5 µm within a 300-µm FOV and <4.5 µm in a 1-mm FOV.