Large-field high-resolution X-ray AKB microscope for measuring hydrodynamic instabilities at the SG-Ⅲ prototype Laser Facility

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

X-ray imaging with a large field of view (FOV) and high resolution is extremely important for Rayleigh–Taylor instability measurement with a small amplitude and high spatial frequency in laser inertial confinement fusion. We developed an advanced Kirkpatrick–Baez (AKB) microscope based on the quadratic-aberration theory to realize a large FOV and high resolution. This microscope was assembled and tested in a laboratory, and it was then successfully applied for imaging the hydrodynamic instability of a perturbation target in implosion experiments at the Shenguang-III prototype laser facility. Imaging results demonstrate that the AKB microscope can achieve an optimal resolution of ~ 0.53μm and ~ 0.40μm and a spatial resolution of <1.5 µm within a 300-µm FOV and <4.5 µm in a 1-mm FOV.