Portable UV-C Device to Treat High Flow of Infectious Aerosols Generated during Clinical Respiratory Care

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Respiratory interventions including noninvasive ventilation, continuous positive airway pressure and high-flow nasal oxygen generated infectious aerosols may increase risk of airborne disease (SARS-CoV-2, influenza virus) transmission to healthcare workers. We developed/tested a prototype portable UV-C 254 device to sterilize high flows of viral-contaminated air from a simulated patient source at airflow rates of up to 100 l/m. Our device consisted of a central quartz tube surrounded 6 high-output UV-C 254 lamps, within a larger cylinder allowing recirculation past the UV-C 254 lamps a second time before exiting the device. Testing was with nebulized A/PR/8/34 (H1N1) influenza virus. RNA extraction and qRT-PCR showed virus transited through the prototype. Turning on varying numbers of lamps controlled the dose of UVC. Viability experiments at low, medium and high (100 l/min) flows of contaminated gas were conducted with 6, 4, 2 and 1 lamp activated (single-pass and recirculation were tested). Our data show 5-log reduction in particle forming units from a single lamp (single- pass and recirculated conditions) at high and low flows. UVC dose at 100 l/m was calculated at 11.6 mJ/cm 2 single pass and 104 mJ/cm 2 recirculated. The protype device shows high efficacy in killing nebulized influenza virus in a high flow of contaminated air.