Determination of Zn (II) on fast-scan anodic stripping voltammetric

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

In this study, the amalgam zinc film on glassy carbon electrode was obtained in situ deposition; it was used to study the determination of Zn (II). Exceptional sensitivity of the determination was demonstrated using fast scan (v < 1 KV/s) anodic stripping voltammetry. The scan rate of 500 V/s was chosen for subsequent determination. In this condition, the calibration curve of Zn (II) was obtained. The concentration of Zn (II) and the peak current showed a good linear relationship from 1×10 − 7 mg / mL to 1×10 − 11 mg / mL for Zn (II). The detection limit attained for Zn (II) was estimated up to 3.33×10 − 12 mg / mL. In addition, Zn (II) was spiked and determined in samples of deionized water. The recovery values for these experiments were between 105.1% and 93.7%, and their relative standard deviation was 3.9%-6.2%. We demonstrate that the proposed method has potential for practical application in analyses of wastewaters and seawaters due to its good anti-interference ability.

Article activity feed