3D Passive Cavitation Mapping (3D-PCM) with a Large-aperture Planar Array

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Urinary stone disease is a common urological condition with increasing incidence, particularly in developed countries. Laser lithotripsy (LL) has become a preferred minimally invasive treatment due to its high precision and low tissue damage. Recent studies suggest that cavitation plays a critical role in stone damage during LL, and three-dimensional passive cavitation mapping (3D-PCM) has emerged as a promising tool for detecting these events. However, clinical translation of 3D-PCM remains challenging due to limitations in imaging depth, field of view (FOV), and procedural compatibility. Here, we present a large-FOV dual-modality imaging system (3D-PCM and B-mode ultrasound) based on a large-aperture planar ultrasound array. Through array optimization and model-based reconstruction, our system achieves an expanded FOV of ~ 40 × 40 mm² at a clinically relevant imaging depth of ~ 110 mm, while maintaining high spatial resolution of ~ 0.6 mm laterally and ~ 0.4 mm axially. In vivo experiments in a porcine model demonstrate that the reconstructed cavitation distribution correlates well with stone damage. Our technology has the potential to provide real-time treatment feedback during LL without disrupting the standard workflow.

Article activity feed