Development and Phantom Validation of a Small-Form-Factor SWIR Emitter Probe for Hydration-Sensitive Spatial-Ratio Measurements in Gelatin-Intralipid Phantoms

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Non-invasive assessment of tissue water content is clinically relevant for edema detection, fluid management, and monitoring of local inflammation. In the short-wave infrared (SWIR), water exhibits strong absorption near 1450 nm with a secondary band near 1650 nm, enabling hydration-sensitive reflectance measurements. However, many SWIR systems rely on spectrometers or high-power broadband sources, limiting translation to compact or wearable platforms. We present a compact SWIR diffuse-reflectance probe built from small-form-factor components using four discrete LEDs (1450 nm and 1650 nm) and a single photodetector to acquire spatially resolved measurements at two source–detector separations (4.5 mm and 7 mm). Probe-geometry-matched Monte Carlo simulations were used to generate lookup tables relating reduced scattering to same-wavelength spatial ratios. 11 A diffusion-based forward model was then used to perform a calibration-anchored water-fraction consistency analysis. Eight gelatin–Intralipid phantoms spanning two scattering conditions and formulation-defined water fractions were evaluated. Spatial-ratio signatures were repeatable and monotonic with nominal water fraction, yielding a mean absolute percent error of 1.55% and a maximum absolute percent error of 3.33% under absorption-consistent conditions. These results demonstrate the feasibility of compact SWIR ratio sensing for controlled hydration changes in tissue-mimicking phantoms and provide a modeling framework for future extension to unknown or in vivo samples.

Article activity feed