Recommendations for sample pooling on the Cepheid GeneXpert® system using the Cepheid Xpert® Xpress SARS-CoV-2 assay

This article has been Reviewed by the following groups

Read the full article

Abstract

The coronavirus disease 2019 (Covid-19) pandemic, caused by SARS-CoV-2, has resulted in a global testing supply shortage. In response, pooled testing has emerged as a promising strategy that can immediately increase testing capacity. In pooled sample testing, multiple samples are combined (or pooled) together and tested as a single unit. If the pool is positive, the individual samples can then be individually tested to identify the positive case(s). Here, we provide support for the adoption of sample pooling with the point-of-care Cepheid Xpert ® Xpress SARS-CoV-2 molecular assay. Corroborating previous findings, the limit of detection of this assay was comparable to laboratory-developed reverse-transcription quantitative PCR SARS-CoV-2 tests, with observed detection below 100 copies/mL. The Xpert ® Xpress assay detected SARS-CoV-2 after samples with minimum viral loads of 461 copies/mL were pooled in groups of six. Based on these data, we recommend the adoption of pooled testing with the Xpert ® Xpress SARS-CoV-2 assay where warranted based on public health needs. The suggested number of samples per pool, or the pooling depth, is unique for each point-of-care testing site and can be determined by the positive test rates. To statistically determine appropriate pooling depth, we have calculated the pooling efficiency for numerous combinations of pool sizes and test rates. This information is included as a supplemental dataset that we encourage public health authorities to use as a guide to make recommendations that will maximize testing capacity and resource conservation.

Article activity feed

  1. SciScore for 10.1101/2020.05.14.097287: (What is this?)

    Please note, not all rigor criteria are appropriate for all manuscripts.

    Table 1: Rigor

    Institutional Review Board Statementnot detected.
    Randomizationnot detected.
    Blindingnot detected.
    Power Analysisnot detected.
    Sex as a biological variablenot detected.
    Cell Line Authenticationnot detected.

    Table 2: Resources

    Experimental Models: Cell Lines
    SentencesResources
    Briefly, virus was cultured in Vero cells in minimum essential media and cellular debris were removed via low-speed centrifugation.
    Vero
    suggested: CLS Cat# 605372/p622_VERO, RRID:CVCL_0059)
    Software and Algorithms
    SentencesResources
    Pooled negative samples were also provided by the Influenza and Respiratory Viruses Program.
    Respiratory Viruses Program
    suggested: (California National Primate Research Center Analytical and Resource Core, RRID:SCR_000696)
    Similar to the statistical approach used by Hanel and Thurner (9), a custom Python script was used to calculate the pooled testing capacity relative to non-pooled testing capacity, represented as a percentage, for each combination.
    Python
    suggested: (IPython, RRID:SCR_001658)

    Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).


    Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.

    Results from TrialIdentifier: No clinical trial numbers were referenced.


    Results from Barzooka: We did not find any issues relating to the usage of bar graphs.


    Results from JetFighter: We did not find any issues relating to colormaps.


    Results from rtransparent:
    • Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
    • No funding statement was detected.
    • No protocol registration statement was detected.

    About SciScore

    SciScore is an automated tool that is designed to assist expert reviewers by finding and presenting formulaic information scattered throughout a paper in a standard, easy to digest format. SciScore checks for the presence and correctness of RRIDs (research resource identifiers), and for rigor criteria such as sex and investigator blinding. For details on the theoretical underpinning of rigor criteria and the tools shown here, including references cited, please follow this link.