Reliability and limits of transport-ventilators to safely ventilate severe patients in special surge situations
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
Background
Intensive Care Units (ICU) have sometimes been overwhelmed by the surge of COVID-19 patients. Extending ICU capacity can be limited by the lack of air and oxygen pressure sources available. Transport ventilators requiring only one O 2 source may be used in such places.
Objective
To evaluate the performances of four transport ventilators and an ICU ventilator in simulated severe respiratory conditions.
Materials and methods
Two pneumatic transport ventilators, (Oxylog 3000, Draeger; Osiris 3, Air Liquide Medical Systems), two turbine transport ventilators (Elisee 350, ResMed; Monnal T60, Air Liquide Medical Systems) and an ICU ventilator (Engström Carestation—GE Healthcare) were evaluated on a Michigan test lung. We tested each ventilator with different set volumes (Vt set = 350, 450, 550 ml) and compliances (20 or 50 ml/cmH 2 O) and a resistance of 15 cmH 2 O/l/s based on values described in COVID-19 Acute Respiratory Distress Syndrome. Volume error (percentage of Vt set ) with P 0.1 of 4 cmH 2 O and trigger delay during assist-control ventilation simulating spontaneous breathing activity with P 0.1 of 4 cmH 2 O and 8 cmH 2 O were measured.
Results
Grouping all conditions, the volume error was 2.9 ± 2.2% for Engström Carestation; 3.6 ± 3.9% for Osiris 3; 2.5 ± 2.1% for Oxylog 3000; 5.4 ± 2.7% for Monnal T60 and 8.8 ± 4.8% for Elisee 350. Grouping all conditions (P 0.1 of 4 cmH 2 O and 8 cmH 2 O), trigger delay was 50 ± 11 ms, 71 ± 8 ms, 132 ± 22 ms, 60 ± 12 and 67 ± 6 ms for Engström Carestation, Osiris 3, Oxylog 3000, Monnal T60 and Elisee 350, respectively.
Conclusions
In surge situations such as COVID-19 pandemic, transport ventilators may be used to accurately control delivered volumes in locations, where only oxygen pressure supply is available. Performances regarding triggering function are acceptable for three out of the four transport ventilators tested.
Article activity feed
-
SciScore for 10.1101/2020.10.07.20208561: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
No key resources detected.
Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We found bar graphs of continuous data. We recommend …
SciScore for 10.1101/2020.10.07.20208561: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
No key resources detected.
Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We found bar graphs of continuous data. We recommend replacing bar graphs with more informative graphics, as many different datasets can lead to the same bar graph. The actual data may suggest different conclusions from the summary statistics. For more information, please see Weissgerber et al (2015).
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-
-