Reducing motion artefact in high resolution 7T MRI using the Magnetic Resonance Minimal Motion (‘MR-MinMo’) head stabilisation device

Read the full article See related articles

Discuss this preprint

Start a discussion What are Sciety discussions?

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

To evaluate the effectiveness of the MR-MinMo head stabilisation device in mitigating motion artifacts during long-duration, high-resolution 7T MRI scans, with and without retrospective motion correction.

Methods

The MR-MinMo was tested on 7 paediatric and 12 adult healthy volunteers using 0.6mm isotropic, 3D Multi-Echo Gradient Echo (ME-GRE) scans. A ∼10-minute ME-GRE scan (linear sampling acceleration factor 2×2) and a ∼20-minute scan (DISORDER sampling acceleration factor 1.4×1.4) were obtained. Both scans were acquired with and without the MR-MinMo in a 2×2 factorial study design. Qualitative and quantitative assessment of image quality of the first echo image was obtained via visual inspection and the normalized gradient squared (NGS) metric respectively. A repeated measures ANOVA was used to determine individual and combined effects of the device and DISORDER images with retrospective motion correction on NGS values, with age as a binary factor (paediatric/adult). T2* maps were generated and the standard deviation of white matter R2*(=1/T2*) values assessed across conditions.

Results

The MR-MinMo significantly reduced motion artifacts both visually and in terms of NGS scores, particularly in paediatric volunteers. There was a significant interaction between MR-MinMo and DISORDER motion correction, suggesting that MR-MinMo improved retrospective motion correction. T2* maps demonstrated improved visual appearance and reduced WM variance with the MR-MinMo.

Conclusion

The MR-MinMo can improve image quality via motion reduction in high-resolution 7T scans By keeping motion within a correctable regime, the device can also improve the performance of retrospective motion correction methods.