Optimal tilt-increment for cryo-ET

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Cryo-electron tomography (cryo-ET) enables high-resolution, three-dimensional imaging of cellular structures in their native, frozen state. However, image quality is limited by a trade-off between angular sampling and radiation damage, making the choice of angular increment during data collection a critical parameter, affecting tomogram quality and downstream analyses. Optimising this increment is challenging due to the high demands on microscope time, storage, and computation. In this study, we systematically evaluated tilt increments of 1°, 2°, 3°, 5°, and 10° using lamellae from Dictyostelium discoideum cells. Keeping total electron dose constant, we found that finer tilt increments (1–3°) produced better-aligned tomograms with higher signal-to-noise ratios and improved outcomes in template matching and subtomogram averaging. A 3° increment emerged as the optimal balance between data quality, alignment accuracy, dose per image, and processing efficiency. This practical recommendation supports both high-throughput and high-resolution structural studies and can guide future cryo-ET data acquisition strategies.