Miniaturized Four-Dimensional Functional Ultrasound for Mapping Human Brain Activity

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Real-time brain monitoring for neurosurgery and neuroscience research of natural behaviors demands portable imaging with high spatiotemporal resolution. Current technologies cannot simultaneously achieve the resolution, mobility, and real-time performance required. Here we present a miniaturized four-dimensional functional ultrasound system capturing volumetric brain hemodynamics in real-time using 3072 transceivers controlled by custom application-specific integrated circuits. The device achieves 450 Hz volumetric imaging up to 8 cm depth while maintaining a form factor suitable for direct cortical placement and potential subcranial implantation. We validated this technology across three clinical scenarios: through skull prosthesis, cranial defect, and during neurosurgery. The system mapped somatotopic finger representations and achieved decoding of individual finger movements during piano playing using machine learning, demonstrating single-trial detection. This portable platform establishes a new approach for brain monitoring bridging laboratory neuroscience and clinical applications, enabling research in natural behavioral settings, providing surgeons real-time hemodynamic feedback, and advancing brain-computer interface development.

Article activity feed