Test-retest reliability of multi-metabolite edited MRS at 3T using PRESS and sLASER

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

Purpose

Spectral editing is the most common MRS approach for noninvasive in vivo measurement of low-concentration, strongly overlapped metabolites in the brain, such as γ-aminobutyric acid (GABA) and glutathione (GSH). Multi-metabolite editing methods, including HERMES and HERCULES, have recently been introduced, where multiple J -coupled metabolites can be edited in a single acquisition without increasing total scan time. Yet little is known regarding the reliability of these methods. This study assessed the test-retest reliability of HERMES and HERCULES, where volume localization was achieved using either PRESS or sLASER.

Methods

Sixteen healthy adult volunteers were scanned twice in two separate sessions. Single-voxel edited MRS data were acquired in the medial parietal lobe using the following sequences: (1) HERMES-PRESS; (2) HERMES-sLASER; (3) HERCULES-PRESS; (4) HERCULES-sLASER. Spectra were processed and metabolites were quantified using the Osprey software. Data quality metrics and reliability statistics were estimated for all four acquisitions.

Results

HERMES-sLASER demonstrated lower within-subjects coefficients of variation (CV ws ) for GSH, glutamine (Gln), and glutamate (Glu) + Gln (Glx), suggesting improved reliability compared to HERMES-PRESS. However, GABA + co-edited macromolecules (GABA+) and Glu showed higher CV ws for HERMES-sLASER. HERCULES-sLASER produced better reliability than HERCULES-PRESS for GABA+, GSH, Glu, Gln, Glx, aspartate (Asp), and lactate (Lac). N -acetylaspartate (NAA) and N -acetylaspartylglutamate (NAAG) showed higher CV ws for HERCULES-sLASER. These findings suggest that sLASER may be more advantageous than PRESS for volume localization in simultaneous multi-metabolite editing.

Conclusion

Using sLASER yielded better test-retest reliability for most metabolites than using PRESS for volume localization for HERMES and HERCULES.