Synchronized EEG with two galvanically-separated miniature wireless behind-the-ear EEG sensors

Read the full article See related articles

Listed in

This article is not in any list yet, why not save it to one of your lists.
Log in to save this article

Abstract

We present a wireless EEG sensor network consisting of two miniature, wireless, behind-the-ear sensor nodes with a size of 2 cm × 3 cm, each containing a 4-channel EEG amplifier and a wireless radio. Each sensor operates independently, each having its own sampling clock, wireless radio, and local reference electrode, with full electrical isolation from the other. The absence of a wire between the two nodes enhances discreetness and flexibility in deployment, improves miniaturization potential, and reduces wire artifacts. A third identical node acts as a USB dongle, which receives and synchronizes the data from the two behind-the-ear nodes. The latter allows to process the 2 × 4 channel EEG as if all 8 channels are sampled synchronously, allowing the use of signal processing algorithms that exploit inter-channel correlations. To demonstrate this synchronized processing, we recorded auditory steady-state responses (ASSRs) at both ears and processed them with data-driven multi-channel filters to optimize the ASSR signal-to-noise ratio, demonstrating a more reliable ASSR detection compared to a single-ear setup.

Article activity feed