Source Reconstruction Without an MRI using Optically Pumped Magnetometer based Magnetoencephalography
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
Source modelling in magnetoencephalography (MEG) maps the spatial origins of electrophysiological signals in the brain. Typically, this requires an anatomical MRI scan of the subject’s head, from which a model of the neuromagnetic field (the forward model) is derived. Wearable MEG – based on optically pumped magnetometers (OPMs) – enables MEG measurement from participants who struggle to cope with conventional scanning environments (e.g. children), enabling study of novel cohorts. However, its value is limited if an MRI scan is still required for source modelling. Here we describe a method of warping template MRIs to 3D structured-light scans of the head, to generate “pseudo-MRIs”. We apply our method using data from 20 participants during a sensory task, measuring induced (beta band) responses and whole-brain functional connectivity. Results show that the group average locations of peak task-induced beta modulation were separated by 2.75 mm, when comparing real- and pseudo-MRI approaches. Group averaged time-frequency spectra were also highly correlated (Pearson correlation 0.99) as were functional connectome matrices (0.87), and global connectivity (0.98). In sum, our results demonstrate that source-localized OPM-MEG data, modelled with and without an individual MRI scan can be similar. This will be useful for future studies where MRI data capture is challenging.