Isobaric labeling update in MaxQuant
Listed in
This article is not in any list yet, why not save it to one of your lists.Abstract
We present an update of the MaxQuant software for isobaric labeling data and evaluate its performance on benchmark datasets. Impurity correction factors can be applied to labels mixing C- and N-type reporter ions, such as TMT Pro. Application to a single-cell multi-species mixture benchmark shows high accuracy of the impurity-corrected results. TMT data recorded with FAIMS separation can be analyzed directly in MaxQuant without splitting the raw data into separate files per FAIMS voltage. Weighted median normalization, is applied to several datasets, including large-scale human body atlas data. In the benchmark datasets the weighted median normalization either removes or strongly reduces the batch effects between different TMT plexes and results in clustering by biology. In datasets including a reference channel, we find that weighted median normalization performs as well or better when the reference channel is ignored and only the sample channel intensities are used, suggesting that the measurement of a reference channel is unnecessary when using weighted median normalization in MaxQuant. We demonstrate that MaxQuant including the weighted median normalization performs well on multi-notch MS3 data, as well as on phosphorylation data. MaxQuant is freely available for any purpose and can be downloaded from https://www.maxquant.org/ .