Direct comparison of venipuncture serum draws versus whole blood finger-stick specimens by anti-COVID-19 IgG/IgM rapid lateral flow immunoassay and ELISA

This article has been Reviewed by the following groups

Read the full article See related articles

Abstract

During the COVID-19 pandemic, manufacturers have developed several diagnostic test kits that include lateral flow immunoassays (LFIA) also known as rapid cassette testing. Rapid cassette testing provides qualitative test results indicating the presence or absence of IgG and IgM antibodies to determine COVID-19 (SARS-CoV-2) infection among individuals. Venipuncture blood draws have been the traditional and widely proposed sample collection method but is costly and not applicable to point-of-care testing (POC) and in remote settings. Whole blood finger-stick blood collections traditionally used by diabetics for glucose level testing is an ideal scenario, but raises concerns regarding the outcome of test results in regards to specificity and sensitivity. In this study we directly compare simultaneous collections of venipuncture serum (SST) blood draws and whole blood finger-sticks (n = 75) to detect human Anti-COVID-19 IgG and IgM antibodies using an EUA-approved lateral flow immunoassay, showing equal to enhanced performance characteristics for this specimen type.

Article activity feed

  1. SciScore for 10.1101/2021.06.04.21258182: (What is this?)

    Please note, not all rigor criteria are appropriate for all manuscripts.

    Table 1: Rigor

    EthicsIRB: 2.1 Study population: This study was approved by the Alcala Pharmaceutical Inc. Institutional Review Board (IORG0010127) in consideration of the Code of Ethics of the World Medical Association (Declaration of Helsinki).
    Sex as a biological variablenot detected.
    Randomizationnot detected.
    Blindingnot detected.
    Power Analysisnot detected.

    Table 2: Resources

    Antibodies
    SentencesResources
    Healgen® lateral flow rapid cassette kits for COVID-19 IgG and IgM antibody detection were used for testing individuals by whole blood finger-stick at the Alcala Labs collection site or San Diego Comprehensive Pain Management Clinic by a certified phlebotomist/nurse.
    IgM
    suggested: None
    The test results indicate the presence of both IgM and IgG anti-SARS-CoV-2 antibodies (Fig 1c).
    anti-SARS-CoV-2
    suggested: None
    For IgG 100 μL HRP-labeled Anti-hIgG tracer antibody was added into the microwells, for IgM 100 μL of HRP-labeled COVID-19 antigen was added.
    Anti-hIgG
    suggested: (Antibodies-Online Cat# ABIN288163, RRID:AB_10795674)

    Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).


    Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.

    Results from TrialIdentifier: No clinical trial numbers were referenced.


    Results from Barzooka: We did not find any issues relating to the usage of bar graphs.


    Results from JetFighter: We did not find any issues relating to colormaps.


    Results from rtransparent:
    • Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
    • Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
    • No protocol registration statement was detected.

    Results from scite Reference Check: We found no unreliable references.


    About SciScore

    SciScore is an automated tool that is designed to assist expert reviewers by finding and presenting formulaic information scattered throughout a paper in a standard, easy to digest format. SciScore checks for the presence and correctness of RRIDs (research resource identifiers), and for rigor criteria such as sex and investigator blinding. For details on the theoretical underpinning of rigor criteria and the tools shown here, including references cited, please follow this link.